S41505: Euglobulin Lysis Time [462n] Print View
Euglobulin Clot Lysis Time   > 60  min 

PRIMARY
1 Plasma Citrated 2 (1) mL   Refrigerated - 4 Hour(s)/Frozen - 21 Day(s)  
ALTERNATE

ECLT provides an overall assessment of the fibrinolysis system by measuring the time for an in vitro clot to dissolve in the absence of the normal plasmin inhibitors. ECLT is useful in assessing the fibrinolytic system and monitoring patients on urokinase or streptokinase fibrinolytic therapy.

Specimen collection container is blue-top (3.2% sodium citrate) tube.
To avoid release of plaminogen activator, do not massage vein
vigorously, pump first excessively or leave tourniquet in place for a
prolonged period. Centrifuge within 30 minutes after collection to get
platelet-poor plasma and freeze on dry ice. Ship specimens frozen on
dry ice. Keep samples in a -60 to -80 degree C freezer if they cannot
be shipped promptly.
Prohibit exercise prior to drawing sample.
Jacobs D., et al, "Laboratory Test Handbook", 4th Edition, Lexi-Comp
Inc., Hudson (Cleveland), 1996, p238.
Setup Schedule
Wednesday

Reported (Analytical Time)
3 days

CPT Code
85360

The CPT codes provided are based on AMA guidelines and are for informational purposes only. CPT coding is the sole responsibility of the billing party. Please direct any questions regarding coding to the payer being billed.

Notes
Methodology: Clot Dissolution






1996 - 2017 Quest Diagnostics Nichols Institute of Valencia